- What Does Genera Scheduling Software Do?
Genera schedules, controls, and monitors automated lab workflows. Genera decides in real time which plate moves where, runs instrument protocols, and connects to your LIMS through APIs.
- Does Genera Work With My Existing Instruments?
Yes. Retisoft writes drivers against each vendor’s API, so Genera controls hardware from many vendors. Retisoft also develops drivers for instruments it hasn’t supported yet.
- Does Genera Run More Than One Process At Once?
Yes. Genera schedules by resource availability, so processes run in parallel. A higher-priority process pauses lower-priority work, runs first, then hands the system back.
- How Do I Estimate Throughput Before Buying An Instrument?
Use estimation. Export real durations from a live run, add the planned instrument to the model, and compare estimated run times for the same workload.
- How Long Does A Genera Trial License Last?
Trial licenses typically run 30 days, and Retisoft extends them when needed. Install the license promptly, because the clock follows a public time server.
Imagine a drug discovery lab on a Tuesday afternoon. The screening team has a plate reader, a washer, a bulk dispenser, a liquid handler, and a collaborative robot arm. Each instrument works reliably on its own. But the scientist running the experiments walks from one bench to another, initiating each protocol manually and staying late to finish the last read.
Nothing in this scenario is inherently wrong. The one obvious improvement is orchestration. A laboratory full of sophisticated equipment doesn’t become an automated laboratory before someone decides which plate goes next, which instrument receives it, and what to do if a step takes longer than expected.
This is where laboratory automation solutions become essential. The right solution doesn’t simply move plates around the lab. It creates a software model of the system that can make decisions, control instruments from any vendor, safeguard delicate incubation steps, keep the show moving if a device becomes unavailable, and notify scientists when human intervention is required.
This article reviews the functions of laboratory automation solutions and explains how Genera – the laboratory workflow scheduling software developed by Retisoft Inc. – helps researchers streamline their processes. Genera can be taught to work with the liquid handlers, plate readers, and robot arms already in the lab and communicates with the LIMS via API. This software takes over the tedious, error-prone parts of an experiment that humans typically handle.
How Genera Turns Existing Lab Instruments Into One Scheduled System
While instruments do the work, orchestration software determines whether it gets done on time. Good laboratory automation solutions such as Genera model the laboratory, control equipment, protect the integrity of time-sensitive assays, and manage incidents. Note that Retisoft Inc. does not manufacture or distribute any readers or liquid handling instruments. Genera turns the devices a laboratory already owns into a fully integrated system.
What A Laboratory Automation Solution Has To Do
Moving a plate from a hotel to a reader is the easy part. The harder work sits around each move. A complete laboratory automation system answers several questions at every step:
- Which plate goes next, and from which shelf?
- The destination nest needs to be empty before the arm arrives.
- Each instrument has to run the right protocol for the plate in front of it.
- When a step fails overnight, the system decides whether the rest of the batch carries on.
- Someone needs to hear about the problem before morning.
Genera handles this through four building blocks: instrument drivers, a workspace model of your system, processes built on a visual canvas, and a scheduler that makes decisions during the run. The sections below follow the same order a new user follows, from installation to a running process. Together, these make up the key features to look for in lab automation scheduling software.
Static Scheduling Versus Dynamic Scheduling
A static schedule determines the exact course of action before a run starts. Each step is assigned a specific place and time within the overall method. The plan is predictable, but it leaves no room for change.
The difference between Genera and other automated scheduling software is that Genera doesn’t try to plan every step. Genera focuses on resource management, which means during the run, Genera is constantly evaluating the available resources. It sorts through the batches, sending plates into the system whenever a resource opens up. Two or more processes move simultaneously, as long as they have access to different instruments.
What Causes Deadlocks in a Fixed Schedule?
A deadlock occurs when a plate waits for a position another plate has seized and will not relinquish. In a fixed schedule, a slight delay in one step, such as an incubation or a plate reader, may throw off every subsequent step, because they were all laid out in advance without considering the possibility of a change. Genera prevents simple deadlocks from occurring by default. Because the software thinks one step ahead, the arm will never carry a plate to a barcode reader that is already in use.
How Decision Nodes Change The Run
A decision node divides a process into a true lane and a false lane. A program, a script, or an instrument command makes the decision. In a Retisoft demonstration, a labware property (reagent type) was set to have values of either Buffer A or Buffer B. Plates marked with values of Buffer A took the true branch to one dispense subprocess; all others took the false branch to another.
Decisions can be taken based upon conditions in the lab. In the same Retisoft demonstration, a process paused waiting for a plate with a read-ready status. The process did not proceed until a user updated the plate status via the software. Once the plate was marked ready, the run proceeded without needing a restart.
What Look Ahead Prevents
The look-ahead function causes Genera to require confirmation beyond the next nest. Adding look-aheads to a series of steps causes the arm to wait until both the barcode reader and the imager are clear before collecting a plate. This will be indicated in the process editor by a bold line across each.
This is the tradeoff to be aware of. Adding look-aheads to every step and Genera acts as a strictly sequential, static scheduler. Adding them only around time-critical steps maintains the throughput of real-time scheduling with the certainty of sequential scheduling where science dictates. Most labs fall somewhere in the middle.
Controlling Instruments From Any Vendor
Most screening labs acquire instruments over several years from various vendors. That’s why Genera was designed to control laboratory hardware from any vendor; the reader you bought in one budget cycle and the washer in the next all work as an integrated laboratory automation system.
One systems integrator at Trent University called Genera a control solution that’s independent of the instrument vendors. Retisoft Inc. doesn’t sell the instruments themselves. They’re valuable as the software that ties them together.
Instrument Driver Packages
Retisoft writes each driver against the instrument vendor’s specific API. Drivers come as package files, often containing drivers for different instruments from the same vendor. Installation requires only a few clicks in the Manage Packages screen, letting users select all packages at once rather than one by one.
Keeping Drivers Current
Vendors update their APIs. A driver written two years ago sometimes falls out of step. Genera lists the version of every installed package, so the first check after an instrument error is simple: confirm the driver is current. The fix for an API change often already exists in a newer package version.
Adding And Swapping Instruments Yourself
Genera lets you configure your own system. In Manage Instruments, select a vendor and model, name the instrument, and define connection parameters such as host name and port. You can change any of these parameters at any time in the instrument control panel, so that you’re not constrained by information that might not be available on day one.
The number of connections on the license defines the number of instruments. Thus, a ten-connection license will accept ten instruments, but will refuse an eleventh. When bringing in a new device, remove one, install the new package, and add the new one.
Modeling Your Laboratory Automation System In Software

Before Genera moves a single plate, the software needs a picture of your bench. Genera calls this picture the workspace, and the workspace has two parts: layouts and projects.
The Layout And The Physical Instruments
A layout represents the physical system using drag-and-drop nodes that represent the robot arm, hotels, stackers, incubators, readers, washers, and dispensers. Each node is linked to an instrument that connects the drawing to the real device.
Some instruments require special handling, and Genera addresses this in its Information tab. For example, a carousel incubator would include a carousel for storage plus a separate nest for the transfer station. A barcode reader becomes a point in space where the arm stops to trigger a scan, then moves on. A liquid handler is represented by a container that holds nothing but the deck nests the arm can reach. Nests contained within the container are automatically associated with the instrument.
The Project, Labware Definitions And Operations
The project defines what goes where. Labware definitions include microplates from 6 to 1536 wells, tip boxes, tube racks, tubes, and lids. Each definition includes a description field for the catalog number, case quantity, and a supplier link. This keeps purchasing details in the software rather than on one scientist’s spreadsheet.
Labware properties turn a plate into a process input. Tag one plate with Protocol C and the reader will run Protocol C on it. Labware operations (read, wash, dispense, incubate, or finished) are user-defined labels that stick to a plate as it moves through the process. A glance at the dashboard instantly reveals where every plate is.
Zones And Instrument Pools
Zones group locations or instruments. Put two hotels in one zone, and Genera picks from whichever hotel holds labware, or fills the first hotel before moving to the second.
Pools matter even more. Suppose three identical washers share a washer zone. A colleague pulls one off the system for a standalone experiment. The run continues on the other two because the process asked for the first available washer, not Washer 1.
Workspaces also travel. The import wizard brings layouts, labware definitions, operations, and zones from another Genera installation, saving you from rebuilding the model during a software update.
Teaching The Robot
Teaching is where laboratory robotics and automation meet the physical bench. Each taught point stores the arm’s X, Y, and Z coordinates along with yaw, pitch, and roll. Get these right once, and the arm repeats them run after run. That repeatability is a big part of how robot automation software reduces error in life science research.
Safe Positions, Approaches And Nests
Every taught location has three kinds of points. Safe positions are spots the arm travels between without hitting any instrument. Approaches are waypoints leading into a nest. The nest is where the plate sits.
Retisoft engineers teach the nest first and work backward. A teach jig in the grippers gives every position the same reference height, which removes variation a regular microplate introduces. To build the approach above a nest, copy the nest values and change only Z. The arm then rises and descends in a straight vertical line.
Mistakes happen, and Genera treats them calmly. During a live teaching session filmed at Retisoft headquarters, the arm clipped the side of a hotel on its way in. The collaborative arm stopped on its soft limit. The engineer cleared the error, switched the move from a joint move to a linear move, and the plate slid straight between the hotel posts. Later in the same session, the arm brushed a thermal cycler beside a liquid handler nest. One extra approach, taught high above the deck, solved the problem.
Why Hotels Only Need Two Shelves Taught
Nobody wants to teach thirteen shelves by hand. In Genera, you teach the bottom and top shelves, and the software interpolates every shelf between them. On a 13-shelf hotel, shelves 1 and 13 are enough. Taught nodes change color in the layout, so untaught positions stand out at a glance.
Testing A Taught Path Before You Run
The test path tool moves the arm through every point, forwards and back (safe, approach, approach, nest) until you see each step is correct. Then just click once to run the full movement in a smooth motion. The grippers open and close from the same screen to test a plate.
You can also pick up labware from the dashboard without writing a process. Right-click on the arm, pick a plate and where you want it, then watch the arm perform the transfer.
You can handle labware differences in a separate table. A thin PCR plate sits lower than the teach jig, so a few-millimeter Z offset will make the arm grip lower for the thin plate only. A labware height value handles stacking and lids. Once you teach the correct positions, they rarely need to change unless an instrument moves or the arm is offloaded and refitted.
Building A Process
A process is where your science lives. Genera’s process editor is a drag-and-drop canvas with two node families. Control nodes shape the flow: decisions, incubations, loops, subprocesses, and labware moves. Activity nodes run commands on instruments. A general category adds message boxes, error triggers, email, and Slack messages.
Lanes And Labware
Each lane follows one piece of labware. A simple liquid transfer, for example, has a sample-plate lane and a destination-plate lane. With lane labware, the user picks the exact plates at a scheduled time from a zone, such as the third hotel of an incubator.
Selecting The Right Labware By Criteria
The alternative is a labware node, which works like a filter. The more criteria you set, the narrower the match: location, labware definition, lidded or unlidded, operational status, barcode, or a custom script returning true or false. Genera looks for plates matching every criterion and ignores the rest.
Decision Nodes, Loops And Subprocesses
Subprocesses package repeated steps, such as adding Buffer A, so the main process stays readable. Loops repeat a block a set number of times, or until a script or instrument command returns false.
Parameter variables add flexibility without duplication. Two scientists share one washing process. One needs five wash cycles, and the other needs ten. Instead of maintaining two versions, the process asks for the number at schedule time.
Linking Lanes So Plates Arrive Together
Genera submits plates to the system as soon as possible. If you have a destination plate that does not yet exist at the start of your method, a sample plate will reach the liquid handler before you have loaded the plate it needs to go to. A loaded tip hovering over an empty slot risks dripping sample out the bottom.
To fix this, right-click on the last destination movement of the plate and select ‘Link move to run protocol’. A dashed line will appear, connecting the two lanes and telling the protocol to wait for both plates to arrive. After the protocol finishes, create a second link to release the plates and let them continue down their separate paths.
Protecting Time-Sensitive Steps
In an ELISA or cell assay, five extra minutes in an incubator can change the data. Real-time scheduling helps throughput, but the science needs guarantees. Genera gives you two.
Incubation Duration and Over-Incubation Tolerance
Each incubation node has an associated duration. Over-incubation tolerance tells Genera how much additional time to allow beyond the duration before it reports an error.
Set over-incubation tolerance to zero if you are using a potentially valuable sample. If the arm is occupied when incubation is due to finish, Genera will abort and report an over-incubation error rather than continuing to run and provide potentially erroneous data.
Reserving The Arm Before An Incubation Ends
The resolution to most over-incubation errors is to configure a look-ahead duration, which reserves the next resource (usually the robot arm) for a specified period.
Setting a five-second look-ahead duration would, for example, ensure the arm held the plate after it left incubation and that no other process could take it during that time.
Running Instrument Checks Inside The Process
Most failed overnight runs fail for dull reasons. A sealer runs out of seals. A labeler runs out of labels. An incubator drifts out of range. Somebody drives back to the lab at 2 a.m. We cover the full list in seven checks to run before leaving an automated run overnight.
Genera 7 introduced the instrument command node. Before Genera 7, a process ran only the activity commands Retisoft judged most useful, and every other command lived in the instrument control panel. Now any command a driver exposes runs inside a process.
In one Retisoft example, a quality control process ran three checks before any samples moved:
- The sealer reported seals remaining, and a decision node flagged any count below five.
- A short script checked the incubator temperature against a range of 30 to 40 degrees. The reading came back at 37, so the run continued.
- The labeler reported three labels left. Genera paused with a message asking the operator to add labels and click OK.
These checks also sit inside decision nodes, loops, and handlers. A start handler, for example, holds the main process until a heated sealer reaches temperature, so no plate meets a cold seal.
You also control how often checks run. A script reading the batch index runs the checks on the first batch only. A parameter variable lets the operator switch quality control on or off at scheduled times. Run the checks on Friday before you leave, and the weekend run starts with full consumables.
Talk to an Automation Engineer
Scheduling, Batches And Priority
Good laboratory workflow scheduling software lets several people share one system without stepping on each other. Genera manages this through batches, parallel processing, and priority.
Batch Counts And Batch Triggers
In Genera, one batch equals one pass through the whole process. A two-lane process with a batch count of five handles ten plates.
Without a batch trigger, Genera fits as many of those batches into the system as resources allow, interleaving plates wherever a nest frees up. Choose the previous batch complete trigger, and Genera finishes one batch before starting the next. On a system with a single reader shared by every process, the trigger keeps the queue orderly.
Running Two Processes At Once
Because Genera schedules by resource availability, two processes always run in parallel if their respective instruments are available. A background workflow and a second user’s workflow run in the same arm, incubator, and reader, without either user having to stop for the other.
Jumping The Queue For An Urgent Sample
Here’s a scene from a Retisoft demonstration. A low-priority imaging workflow runs on repeat, pulling plates from ambient storage. A colleague walks in with a precious sample that can’t sit at room temperature. In most labs, this means the person pulls the imager offline and feeds the plates by hand.
In Genera, the colleague schedules the urgent workflow at priority 10 and starts it right up. Genera finishes the current plate, pauses the background workflow, runs every urgent batch in order, and automatically resumes the background workflow. Batch variables keep track of where every plate came from, so every plate goes back home. Priority ranges up to 10, and three processes at the same priority run in order.
Recurring And Scheduled Work
A process starts immediately, waits for user input, or starts at a set day and time. Recurrence repeats a process like a calendar invite: when finished, every minute, hourly, daily, or weekly. The repeats end after a set number of runs, at a date and time, or never.
Take an absorbance reading every hour until next Friday at 8:30 a.m., when the assay ends. Or a quality control process every Monday morning and Friday night. Genera shows every scheduled run in day, week, and month calendar views.
One honest limit: recurrences persist in the database if Genera closes, but Genera has to be open when a run is due. A run moving plates when the software closes is lost, though the remaining schedule survives.
Simulation, Estimation And Modeling Throughput
The most challenging aspects of the automation business case involve demonstrating benefits before procuring the equipment. Genera software provides two options to help you do that: simulation and estimation. Simulation employs a process flow with instruments running in a simulated mode and produces a visual display of plates being transported in the defined workspace. Estimation uses the same process flow and, using either mean or max values for activity durations, calculates the overall run time without a visual display. Genera will cycle through the number of batches and return an estimated total run time based on the activity durations within seconds.
The durations can come from two sources: they can be entered manually at the activity level, or imported from a completed live run using either the mean or max time per activity across the batches.
A webinar on ELISA included a demonstration that highlighted the value of the estimation feature. Retisoft partner Lab Integrators, a division of Copia Scientific, designed and built the workcell used in the webinar, which included nine instruments. After a live run of the process, Retisoft’s applications engineer imported the actual durations from that run and estimated the time for running ten plates with ten wash cycles per plate. With a single washer instrument, the estimate was 2 hours 32 minutes; with a second washer added, the same workload had an estimated run time of 1 hour 41 minutes. This time difference would factor into a capital request justification and was developed before the purchase order was issued.
Handling Errors Without Scrapping A Run
An error at batch seven of ten should not cost you batches eight, nine, and ten. When an activity fails, Genera offers five choices: retry the activity, ignore the error and continue, stop the batch, stop the process, or reschedule the whole run.
In the process creation walkthrough, a reader refused a plate without a barcode. The operator typed a barcode into the software, clicked Retry, and the run continued. With ten batches running, a single faulty plate gets stopped on its own while the other nine finish.
During a run, the task view lets you pause, end early, or add batches. A batch chart, a log file, and live variable values show exactly where each plate stands.
Notifying People When Something Happens
Handlers run a program, script, or instrument command. Common handler triggers are start, finish, and error.
When an error occurs, Genera can send a message by email, Microsoft Teams, or Slack. Teams and Slack messages are sent via a channel webhook so that everyone on the channel sees the alert rather than a select group by email. Every error message logs the time, process, activity, batch, labware, and error text. A scientist working from home can see if they need to rush in and resolve the issue or wait until the morning.
Handlers also copy between processes. You can create a start handler to initialize every instrument and a finish handler to turn them all off once, then paste them into every new process. The Teams and Slack programs are not installed by default and are provided by Retisoft support on request.
Managing Users And Protecting A Validated Process
Once a process is validated, it must remain as validated. Genera includes user management as a basic platform capability. Turn on the feature, and each person logs in with a role.
The default roles are admin, integrator, operator, and guest. An integrator builds processes and configures instruments. An operator opens, schedules, and runs processes and initializes instruments, but can’t edit a process or change instrument settings. A guest only views.
Admins can also create custom roles, such as a scientist role that can schedule a process and nothing else. Fewer menus, less confusion, and the gold standard process stays untouched.
Scripting When You Need It, And When You Do Not
Most processes need no code. Drag, drop, and fill in parameters.
When a step needs more, Genera offers two levels. An expression is one line inside a parameter, such as a message reading Batch 2 of 3. A script node runs several lines, such as the temperature range check described earlier. Built-in variables expose the batch index, batch count, loop index, the current plate, and the full run context. API documentation ships with every Genera installation, and comments inside a script leave notes for the next person.
One deliberate limit: scripts run commands but do not define new classes. Our advice is to start without code and add a script only when a parameter falls short of the rule.
Connecting Genera To Other Systems
Integrated laboratory automation rarely stops at the instruments. Genera connects to your LIMS and other software through APIs. API access is a hallmark of lab automation platforms that integrate with existing equipment.
The Remote API plugin, installed as a package, accepts RESTful commands and follows the same user management rules as the desktop software. Through the API, an outside system initializes instruments, adds labware with its definition, location, barcode, and properties, and launches a process with a batch count, a priority, and per-batch parameters. In one demonstration, plate one received Protocol 1 and plate two received Protocol 2, both set through the API. The system then polls the task until it finishes and clears the labware for the next run.
Consider a LIMS sending tonight’s plate list and the reader protocol for each plate. Genera runs the work, and the LIMS checks progress without anyone touching the scheduler. Some Retisoft customers already run their own internal interfaces on top of Genera this way.
For simpler data exchange, a built-in activity reads a CSV file into process variables, so a worklist decides which protocol each plate receives.
Licensing, Device Connections And Getting Started
Getting Genera running takes less ceremony than most lab software. Retisoft sends a folder with the installer, your license, and your driver packages. Double-click the installer, accept the defaults, load the license, and install the packages. Your IT team only steps in if the computer requires admin rights.
Three license details are worth knowing:
- Trial licenses typically run for 30 days, and Retisoft can extend them as needed. The clock follows a public time server, not the computer’s date, so install the license the day it arrives.
- A permanent license is tied to one PC’s unique ID and does not transfer.
- Each license specifies an edition (Standard, Pro, or Lite) and several device connections, which caps how many instruments you can add.
The Retisoft website offers a free, fully functional trial of Genera. For simpler applications such as plate reading and washing, liquid handling, labeling, or dispensing, GENERA Lite is a compact, self-deployed option delivering walk-away time without a complex integration project. Labs planning to grow use X-LAB, Retisoft’s modular table system, where tables are added without redoing earlier work.
Lab Automation Services, Training And Ongoing Support
Installation day is where your relationship with Retisoft begins, not where it ends. Our lab automation services include engineering and scientific application consulting, custom driver development, software support, and hardware and maintenance support.
The Retisoft Hub delivers on-demand training, support, user forums, and product updates. Learn Genera Gems, a video series that showcases tips and techniques from existing Genera users about features such as recurrence, scripting, and the Remote API. Viewers can suggest topics for future videos.
Hear what customers are saying about the value of Retisoft training. “After a week of training, I was able to start writing my own drivers, plugins, and custom web pages,” says a senior research engineer at the Johns Hopkins University School of Medicine. “We built a system that includes a robot arm, liquid handler, incubator, and readers.” The director of lab automation at Pfizer notes, “We brought all our robotic platforms and tools up and running within six months.” When pandemic restrictions limited travel in 2020, Retisoft provided the expertise and support to help automate and implement COVID-19 testing and research through online training and remote system support.
Why Laboratories Work With Retisoft
Retisoft Inc. has focused on lab automation scheduling since 1998, from its base in Mississauga, Ontario. Genera systems run in North America, Europe, and Asia, through a network of partners Retisoft calls automationships, including integrators such as Copia Scientific.
Labs cite a few key reasons. Genera is vendor-independent, so you keep the instruments you trust. Real-time scheduling keeps shared systems moving. Users configure, model, and teach their own systems. And when something needs a new driver or a package not shipped by default, Retisoft builds or supplies it. It’s the core difference between automation-focused and equipment-centric scheduling platforms.
The best way to judge fit is to see your own workflow on screen. Book a Genera demo, bring your instrument list, and ask the team to model your bottleneck.
Map Your Workflow With Our Team
From A Room Full Of Instruments To An Automated Lab
Go back to the screening team from the start of this guide. Same reader, same washer, same liquid handler, same robot arm. Nothing new on the bench. The difference sits in the software deciding what happens next.
With Genera scheduling the work, plates leave the incubator on time. A colleague’s urgent sample jumps the queue without anyone pulling an instrument offline. When a labeler runs low at 6 p.m. on a Friday, the run pauses before samples move, not halfway through the night. The scientist who used to stay late for the final read goes home and checks a Slack message instead.
None of this asks you to replace equipment you trust. Retisoft Inc. builds the scheduling layer, and your instruments stay yours.
If your lab sits somewhere between a few connected instruments and a fully integrated system, start small. Model one bottleneck, run an estimate, and see what the numbers say. The Genera team is happy to walk through your own workflow in a demo whenever you’re ready.